Transfer of calibrations between projects

QTM automatically uses the latest calibration performed on the computer, even if it has been made in another project using the same camera system. This ensures that you always use the most up-to-date calibration for your measurements.

The calibration is also checked if the camera configuration has changed when locating the camera system or starting the cameras in preview mode. If the latest calibration performed on the computer matches the new camera configuration, it will be loaded into the project.

More precisely, the latest calibration is transferred when the project contains the same camera linearization files. The order of the cameras in the project does not matter.

Vice versa, the latest calibration will not be transferred when the project contains different linearization files, even if the camera system is the same. This is for example the case when the latest calibration is an advanced calibration.

This has the following implications:

  • If you want to use the latest advanced linearization in another project, you will need to manually load it.

  • If you want to avoid that calibrations transfer between projects, make sure that the projects do not share the same linearization files, for example by using the advanced calibration method.

  • If you want to use optimized linearizations for two or more projects and want them to automatically use the latest calibration, make sure to manually load the same advanced calibration into these projects. Then, for new calibrations, use the standard calibration method using the linearizations loaded in the project, see chapter Choice of linearizations.