Specification of force plate location
The specification of the force plate location is needed to map the forces measured in the local coordinates of the force plates onto the global coordinate system of QTM.
For most force plates, the Z-axis is pointing down, and the Y-axis is pointing in the direction of the long side of the force plate, with the positive being in the direction that points away from the connectors of the force plate. For detailed information, see the specifications from the manufacturer.
The force plate location can be generated from a capture with markers placed in the corners of the force plate, or specified manually.
Generate force plate location from a capture
For automatic generation of the force plate location, place markers in the four corners of the force plate.
The markers do not need to be placed on the exact top of the corners, but it is important that they are placed symmetrically in the four corners for a correct estimation of the center of the force plate.
Follow these steps for automatic generation of the force plate location. If you have multiple force plates, these steps must be repeated for each individual force plate.
-
Make a short capture with the markers placed in the corners of the force plate.
-
Identify the four markers with a label of your choice, and keep the capture file open.
-
Open the Force plate page and click Generate. QTM tries to identify the corners of the force plate by comparing them to the width and length of the plate (as entered in the Force plate calibration parameters dialog). When QTM finds solutions, the following dialog is displayed:
-
Click OK to open Load measured force plate location dialog, see below, and select one of the solutions found by QTM. The local coordinate system of the force plate is shown for reference.
If you are unsure which solution to choose, just pick one. The orientation can be easily changed later in the Force plate location dialog via the View/Edit button.
-
The vertical level of the force plate will be in the plane of the marker centers. To compensate for the height offset, use the Adjust for marker offset setting in the Load measured force plate location dialog. You can also click View/Edit in the Force plate settings dialog and manually adjust the vertical position.
-
Make sure that the orientation is correct, following the instructions in chapter Checking the position and orientation of the force plate.
Manual revision and specification of force plate location
To manually specify the location, click View/Edit, and enter the global X, Y and Z coordinates (in mm) of the four corners of the force plate according to the orientation in which it is mounted. The local coordinate system of the force plate is shown for reference.
Make sure that the orientation is correct, following the instructions in chapter Checking the position and orientation of the force plate. Use the internal coordinate system of the force plate, shown in the dialog, to determine the orientation of the force plate corners.
Checking the position and orientation of the force plate
It is recommended to make a test measurement after the location process to make sure that the position and orientation of the force plate are correctly mapped.
-
Make sure that the force plate is properly zeroed before testing.
-
Enable display of force vectors to show the ground reaction force (GRF) in the 3D view of QTM.
-
Start a preview in QTM or make a short capture.
-
Apply a load to the force plate, for example with your foot with a marker, probing different locations.
-
Inspect the force vector in the 3D view.
-
If the origin of the force vector is at the opposite side of the plate, you can use the Rotate 180 degrees option in the Force plate location dialog, shown above.
-
If the force vector is pointing down, the force plate location must be flipped by mirroring the corner assignments either left-right or top-down.
-
For any other position errors of the force vector, you will need to redo the force plate location entirely.
-
Validity of the force plate location
The force plate location is valid for the calibration for which it has been defined. Using the same location definition for another calibration requires that the L-frame is placed in the exact same location, for example by using an L-frame with positioning plates, see chapter Qualisys calibration kits.
Notes on the validity of force plate location:
-
Force plate locations are not automatically adapted when you add a transform to the calibration. In that case, you will need to redefine the force plate locations.
-
For the highest accuracy of the force plate location, it is recommended to redefine the location for each individual calibration.